PREL201908772
There are no deposits or bonds for this permit application.
| Inspection | Outcome | Requested | Scheduled | Date Complete |
|---|---|---|---|---|
| Electrical Final Inspection | Pending | |||
|
Corrections:
Miscellaneous/Other
Outstanding
10/10/2019
No sign installed. Sign installation will require separate permit.
|
||||
|
Miscellaneous/Other
Outstanding
10/10/2019
Verify if RTU is existing or new install. 50a OCP with number 10 wires, verify nameplate requirements.
|
||||
| Electrical Rough-in Inspection | Failed | 8/6/2019 (12:30 - 2:30) | 8/6/2019 (PM) | 08/06/2019 |
|
Corrections:
Miscellaneous/Other
Carried Forward
08/06/2019
Patient care areas require redundant grounding systems per NEC 517.
Adjustment rooms and possibly consult rooms.
|
||||
| Electrical Rough-in Inspection | Partial Pass | 8/8/2019 (10:00 - 12:00) | 8/8/2019 (AM) | 08/08/2019 |
|
Corrections:
Miscellaneous/Other
Resolved
08/08/2019
Patient care areas require redundant grounding systems per NEC 517.
Adjustment rooms and possibly consult rooms.
|
||||
|
Miscellaneous/Other
Comment Only
08/08/2019
EMT installed for patient areas wall RI- ok
|
||||
| Electrical Rough-in Inspection | Passed w/ Conditions | 10/1/2019 (2:30 - 3:30) | 10/1/2019 (PM) | 10/01/2019 |
|
Corrections:
Miscellaneous/Other
Comment Only
10/01/2019
Above ceiling - ok
Couple of Jboxes left to splice and cover.
|
||||
|
Miscellaneous/Other
Comment Only
10/01/2019
Left note for General permits required for all FA work and Temp control work.
|
||||
|
Miscellaneous/Other
Comment Only
10/01/2019
Verifying listing and wiring requirements of Xray machine
|
||||
| Electrical Final Inspection | Failed | 10/7/2019 (12:30 - 2:30) | 10/7/2019 (PM) | 10/07/2019 |
|
Corrections:
Miscellaneous/Other
Carried Forward
10/07/2019
In the procedure rooms the isolated ground receptacles are not allowed. per NEC 517.16A
Receptacles need to be hospital grade with the green dot.
|
||||
|
Miscellaneous/Other
Carried Forward
10/07/2019
Metal boxes must be connected to EGC in all areas including procedure rooms and be connected to the receptacles.
|
||||
|
Miscellaneous/Other
Carried Forward
10/07/2019
EGC in panel and disconnect to Xray machine have doubled up wires on lugs. Lugs do not appear to be rated for 2 wires.
|
||||
|
Miscellaneous/Other
Carried Forward
10/07/2019
No sign installed
|
||||
|
Miscellaneous/Other
Carried Forward
10/07/2019
Verify if RTU is existing or new install.
|
||||
| Electrical Final Inspection | Partial Pass | 10/10/2019 (8:00 - 10:00) | 10/10/2019 (AM) | 10/10/2019 |
|
Corrections:
Miscellaneous/Other
Resolved
10/10/2019
In the procedure rooms the isolated ground receptacles are not allowed. per NEC 517.16A
Receptacles need to be hospital grade with the green dot.
|
||||
|
Miscellaneous/Other
Resolved
10/10/2019
Metal boxes must be connected to EGC in all areas including procedure rooms and be connected to the receptacles.
|
||||
|
Miscellaneous/Other
Resolved
10/10/2019
EGC in panel and disconnect to Xray machine have doubled up wires on lugs. Lugs do not appear to be rated for 2 wires.
|
||||
|
Miscellaneous/Other
Carried Forward
10/10/2019
No sign installed. Sign installation will require separate permit.
|
||||
|
Miscellaneous/Other
Carried Forward
10/10/2019
Verify if RTU is existing or new install. 50a OCP with number 10 wires, verify nameplate requirements.
|
||||